Endress+Hauser CCM 223 Chlorine Analyzer (w/ pH Compensation), Panel Mount, CCM223-EP1105
Description
- Alarm contact and two analog outputs
- Continuously monitor either free or total chlorine residual
- Features pH compensation
Endress+Hauser CCM223 chlorine analyzers continuously monitor free or total chlorine in effluent or potable water. pH compensation lets you maximize your disinfecting process in free chlorine applications and is highly encouraged. The CCM223 chlorine analyzer features a 1/4 DIN housing for mounting in an existing control panel.
The CCM223 chlorine analyzer features a 1/4 DIN housing for mounting in an existing control panel.
Analyzers with optional HART®, PROFIBUS® PA or PROFIBUS DP output communications are also available; contact us for more information.
- Accuracy: ±0.5% of reading
- Display: backlit LCD
- Sample flow rate: =0.133 gpm
- Analog outputs: (2) isolated 0-20 or 4-20 mA
- Alarm relay: 1 SPDT rated 2A @ 250 VAC
- Temperature compensation: automatic, 32 to 130°F
- Sensor to analyzer: 15'L cable (included)
- Power: 115 VAC, 50/60 Hz (alternate voltages available)
- Enclosure: none (panel mount)
- Approvals: CE, CSA
Product Information
- Item number
- 64029
- Brand Name
- Endress and Hauser
- Catalog Page
- 1093
- Net weight
- 1
- Condition
- New
Endress+Hauser CCM 223 Chlorine Analyzer (w/ pH Compensation), Panel Mount, CCM223-EP1105
- Item number
- 64029
- Brand Name
- Endress and Hauser
- Catalog Page
- 1093
- Net weight
- 1
- Condition
- New
- Alarm contact and two analog outputs
- Continuously monitor either free or total chlorine residual
- Features pH compensation
Endress+Hauser CCM223 chlorine analyzers continuously monitor free or total chlorine in effluent or potable water. pH compensation lets you maximize your disinfecting process in free chlorine applications and is highly encouraged. The CCM223 chlorine analyzer features a 1/4 DIN housing for mounting in an existing control panel.
The CCM223 chlorine analyzer features a 1/4 DIN housing for mounting in an existing control panel.
Analyzers with optional HART®, PROFIBUS® PA or PROFIBUS DP output communications are also available; contact us for more information.
- Accuracy: ±0.5% of reading
- Display: backlit LCD
- Sample flow rate: =0.133 gpm
- Analog outputs: (2) isolated 0-20 or 4-20 mA
- Alarm relay: 1 SPDT rated 2A @ 250 VAC
- Temperature compensation: automatic, 32 to 130°F
- Sensor to analyzer: 15'L cable (included)
- Power: 115 VAC, 50/60 Hz (alternate voltages available)
- Enclosure: none (panel mount)
- Approvals: CE, CSA