GF Signet 9950 Transmitter; Panel Mount, AC/DC Powered, 3-9950-2
Description
- Large auto-sensing backlit display offers highest visibility
Signet's 9950 universal-input transmitters provide a dual-channel interface for a wide variety of parameters including pH/ORP, conductivity/resistivity, salinity, flow, pressure, temperature and level. Even 3rd party generic 4-20mA sensor signals can be displayed with the use of i-Go signal converter (sold separately). The large, auto-sensing display is readable at 4 to 5 times the distance of traditional transmitter displays, reducing the risk of misreading values.
The dual-channel 9950 transmitter simultaneously displays a bar graph and up to two process values for each channel (i.e., both temperature and pH). This transmitter offers advanced features such as derived functions (control of relays or analog loop with sum, difference, or ratio of the two inputs). An onboard clock & calendar let you schedule routine relay actions, such as closing a relay to perform purge cleaning on Tuesday at 10 am. Boolean logic functions allow up to three signals to control a single output relay; for example, if input signal A and B are both hi-alarm, close the relay output. Add one of three optional plug-in modules for even greater process control. Choose from modules with relay outputs, solid state switch outputs, and binary (digital) inputs.
9950 transmitters are available for panel mounting only. Use with any standard 1/4 DIN enclosure, such as stock # 67727.
For a complete system, order transmitter, sensor (ph, ORP, Conductivity or flow), sensor electronics or preamplifier, sensor installation fittings for inline flow-through applications, and mounting kit (for field mounting) or enclosure (for panel mounting).
- Input types: Digital: S3L (serial ASCII, TTL, 9600 bps); Frequency: 0.5 to 1500 Hz; accuracy 0.5% of reading
- Range: -1.00 to 15.00 pH; -1999 to 1999.9 mV; -9999 to 99,999 units per second, minute, hour or day (flow rate); 0.000 to 99,999 µS, mS, ppm, ppb, kΩ, MΩ (conductivity, TDS, resistivity)
- Outputs: Analog: standard (1) passive 4-20 mA in base unit; optional (1) additional via module (sold separately); Relay: optional (2) dry contacts via module (sold separately), 5A max; Open Collector: standard (1) in base unit, max 30 VDC
- Operating conditions: 14 to 140°F
- Power: Loop or 10.8 to 35.2 VDC (24 VDC nominal); 100 to 240 VAC, 50-60 Hz
Product Information
- Item number
- 21273
- Brand Name
- GF Signet
- Catalog Page
- 1129
- Net weight
- 2
- Condition
- New
Accessories
Related
GF Signet 9950 Transmitter; Panel Mount, AC/DC Powered, 3-9950-2
- Item number
- 21273
- Brand Name
- GF Signet
- Catalog Page
- 1129
- Net weight
- 2
- Condition
- New
- Large auto-sensing backlit display offers highest visibility
Signet's 9950 universal-input transmitters provide a dual-channel interface for a wide variety of parameters including pH/ORP, conductivity/resistivity, salinity, flow, pressure, temperature and level. Even 3rd party generic 4-20mA sensor signals can be displayed with the use of i-Go signal converter (sold separately). The large, auto-sensing display is readable at 4 to 5 times the distance of traditional transmitter displays, reducing the risk of misreading values.
The dual-channel 9950 transmitter simultaneously displays a bar graph and up to two process values for each channel (i.e., both temperature and pH). This transmitter offers advanced features such as derived functions (control of relays or analog loop with sum, difference, or ratio of the two inputs). An onboard clock & calendar let you schedule routine relay actions, such as closing a relay to perform purge cleaning on Tuesday at 10 am. Boolean logic functions allow up to three signals to control a single output relay; for example, if input signal A and B are both hi-alarm, close the relay output. Add one of three optional plug-in modules for even greater process control. Choose from modules with relay outputs, solid state switch outputs, and binary (digital) inputs.
9950 transmitters are available for panel mounting only. Use with any standard 1/4 DIN enclosure, such as stock # 67727.
For a complete system, order transmitter, sensor (ph, ORP, Conductivity or flow), sensor electronics or preamplifier, sensor installation fittings for inline flow-through applications, and mounting kit (for field mounting) or enclosure (for panel mounting).
- Input types: Digital: S3L (serial ASCII, TTL, 9600 bps); Frequency: 0.5 to 1500 Hz; accuracy 0.5% of reading
- Range: -1.00 to 15.00 pH; -1999 to 1999.9 mV; -9999 to 99,999 units per second, minute, hour or day (flow rate); 0.000 to 99,999 µS, mS, ppm, ppb, kΩ, MΩ (conductivity, TDS, resistivity)
- Outputs: Analog: standard (1) passive 4-20 mA in base unit; optional (1) additional via module (sold separately); Relay: optional (2) dry contacts via module (sold separately), 5A max; Open Collector: standard (1) in base unit, max 30 VDC
- Operating conditions: 14 to 140°F
- Power: Loop or 10.8 to 35.2 VDC (24 VDC nominal); 100 to 240 VAC, 50-60 Hz